NEWFIRM Manual
Here we describe the NEWFIRM observing GUI and discuss other information relevant for observing.
- Startup
- NEWFIRM Observing GUI
- NEWFIRM IRAF Interface
- Daytime Calibrations
- Telescope Pointing
- Telescope Focus
- Observing
- Shutdown
Startup:
NEWFIRM Observing GUI
- Open VNC viewers [NEWFIRM Obs and NEWFIRM IRAF and NEWFIRM NICC (optional)]
- The NEWFIRM support scientist of Telescope Operations support staff will help
- Start NEWFIRM GUI in Obs VNC viewer
- Press "Start of Night"
- This should open the environmental cover.

- Check that environmental cover is open on the NEWFIRM NICC vncviewer


- Turn on detector power


- Open filter tool

- Initialize filters
- Move filter position to "DARK" if the dark filter is not already in place
- Set Proposal ID (PropID)
- This will load the proposal PI and co-Is
- No images can be taken unless the PropID is set to a valid PropID
- Take a junk image to reset image counter/number
- The image counter does not reset to 1 at the beginning of the night
- Make sure dark filter is in place
- Under Exposure Control:
- Set Basename to "ajunk_"
- Set Title to "Junk"
- Set Exp. Time to "1.120"
- Set Repeats to "0"
- Set Coadds to "0"
- Set Fowler to "1"
- Set Sequence to "1"; This must be the last step.
- Click "Expose"
- The image counter will now be reset and calibration data will start with image number *0002*
NEWFIRM IRAF Interface
Daytime Calibrations
Telescope Pointing
Telescope Focus
Observing
Shutdown
Updated on May 14, 2025, 1:08 pm