NEWFIRM Manual

Here we describe the NEWFIRM observing GUI and discuss other information relevant for observing.

  • Startup
    • NEWFIRM Observing GUI
    • NEWFIRM IRAF Interface
  • Daytime Calibrations
  • Telescope Pointing
  • Telescope Focus
  • Observing
  • Shutdown

Startup:

NEWFIRM Observing GUI

  • Open VNC viewers [NEWFIRM Obs and NEWFIRM IRAF and NEWFIRM NICC (optional)]
    • The NEWFIRM support scientist of Telescope Operations support staff will help
  • Start NEWFIRM GUI in Obs VNC viewer

Start NEWFIRM GUI Start NEWFIRM GUI

  • Press "Start of Night"
    • This should open the environmental cover.
NEWFIRM Start of Night
  • Check that environmental cover is open on the NEWFIRM NICC vncviewer
NEWFIRM Environmental Cover "Closed" NEWFIRM Environmental Cover "Open"
  • Turn on detector power
NEWFIRM Detectors "Off" NEWFIRM Detectors "On"
  • Open filter tool
NEWFIRM Filter Tool
  • Initialize filters
  • Move filter position to "DARK" if the dark filter is not already in place
  • Set Proposal ID (PropID)
    • This will load the proposal PI and co-Is
    • No images can be taken unless the PropID is set to a valid PropID
  • Take a junk image to reset image counter/number
    • The image counter does not reset to 1 at the beginning of the night
    • Make sure dark filter is in place
    • Under Exposure Control:
      • Set Basename to "ajunk_"
      • Set Title to "Junk"
      • Set Exp. Time to "1.120"
      • Set Repeats to "0"
      • Set Coadds to "0"
      • Set Fowler to "1"
      • Set Sequence to "1"; This must be the last step.
      • Click "Expose"
      • The image counter will now be reset and calibration data will start with image number *0002*

NEWFIRM IRAF Interface

Daytime Calibrations

Telescope Pointing

Telescope Focus

Observing

Shutdown